This makes signals with a significant frequency modulation look very bad on the screen. Properly choosing the span this inconvenience is soslayable.

nanovna spectrum analyzer

Recognized, the signal must fall exactly on any of the sampled frequencies because otherwise it is strongly or completely attenuated. The third problem I have appreciated is that to be Of the two peaks, the "real" is the higher frequency (the one on the right). In a "real" spectrum analyzer this image frequency is separated at several GHz and is rejected by the input LPF. Secondly, below 1MHz of span two peaks appear instead of one, as the image frequency also appears separated twice the value of the IF (i.e. Known that offset, is not a major problem. I assume that it is because the nanoVNA accepts as 0dB the value applied in its CH1 in the calibration, which is that provided by CH0 and has those values ​​(in the measurement frequency). The second thing I am going to say is that I have been trying to check the operation of the CH1 input as a spectrum analyzer and I have been able see under what circumstances it is more or less usable.įirst, the level measurements appear to be around 15 dB higher than the signal applied below 300 MHz and around 20 dB higher to over this frequency. I received my nanoVNA a couple of weeks ago and I have been doing many tests with it.įirst I must say my unit came without shields and I added them, but there was no considerable improvement in the base noise level. Hello again Eric and other possible readers.